  1. The HM1005 is a multi-function oscilloscope loaded with features such as 
  2. true 3-channel operation, a genuine second time base, and even a separate 
  3. second trigger facility.  
  4. Up to 6 traces can be displayed for evaluating waveform relationships by 
  5. using the alternating time base mode and all three input channels.  In this
  6. configuration, the normal signal is displayed together with the expanded 
  7. signal section.  The second trigger facility, with its own separate slope 
  8. and level control, provides for the stable viewing of normally troublesome 
  9. asynchronous signal components.  
  10. The main time base trigger circuit is designed to provide reliable 
  11. triggering to greater than 130 MHz, at signal levels as low as 1 division. 
  12. A user-selectable x10 magnifier extends the maximum sweep range to 
  13. 5ns/division. 
  14. An active TV Sync Separator extracts clean field and line sync. pulses to 
  15. ensure a high degree of display stability when viewing composite video 
  16. signals. The built-in delay line permits viewing of the trigger edge at 
  17. all times. The overscan feature indicates if any part of the trace passes 
  18. beyond the vertical limits of the CRT screen.  
  19. An analog Y-output, switchable to Channel I or II, allows further external 
  20. processing of the signal.  Another new feature on the HM 1005 is the easy to
  21. read 3-digit LED delay time multiplier for accurate measurements. 
  22. The HM 1005 offers the right combination of triggering control, frequency 
  23. response and time base versatility, to facilitate measurements in every 
  24. enviroment from Workshop to Field service. 
  25.  
  26. Specifications
  27.  
  28. Vertical Deflection
  29. Operating Modes:        Ch.1 or Ch.11 individually.
  30.                         Ch.1 or 11 alternate or chopped(ch/freq.0.5MHz)
  31.                         Ch.111 can be displayed with Ch.1 and 11
  32.  
  33. Sum or Difference:    From Ch.1 and 11 
  34.                         (with dedicated "invert" control for Ch.11)
  35. X-Y Mode:               Via Channel 1(X) and Channel 11(Y)
  36. Frequency Range:        3 x DC to 100MHz (▒3dB).
  37. Risetime:               <3.5nS.  Overshoot: <1%
  38. Deflection Coefficient(Ch.1/11):    
  39.                         10 calibrated steps from 5mV to 5V/div ▒3% 
  40.             in 1-2-5 sequence, variable 2.5:1 to 
  41.                         min.12.5V/div
  42. Y Mag. x5:              Calibrated to 1mV/div ▒5%(Freq DC to 10MHz)
  43. Deflection Coefficient(Ch 111):    50mV/div.▒3% variable to min 250mV/div
  44. Input Impedance Ch.1 -111:    1MR # 22pF
  45. Input Coupling(Ch.1 - 111):    DC-AC GD(Ground)
  46. Input Voltage:        Max 400V(DC ▒ peak AC)
  47. Y Output from Ch. or 11:    ~45mV/div (50R). Delay line: approx. 90nS
  48. Triggering:             With automatic 10Hz-120MHz(>5mm height)
  49.             normal, DC-130MHz with"Level"adjust(>10mm)
  50. Slope:            Positive or Negative. LED indication for trigger
  51. Coupling:            AC (>10hZ-40MHz),  DC (0-40MHz),
  52.             HF (15kHz-130MHz),  LF (0-1kHz)
  53. TV Sync-seperator:      For line and frame, pos / neg
  54. Triggering Timebase B:    (Internal only) with level control
  55.             (AC coupling) slope positive or negative
  56.  
  57. Horizontal Deflection   
  58. Timebase A:        23 calibrated steps from 50nS to 1S/div.▒3%
  59.             in1-2-5 sequence, variable2.5:1 to min 2.5S/div
  60.             with Mag. x 10 to 5nS/div ▒5%
  61. Timebase B:        21 calibr. steps from 50nS to0.2S/div.1-2-5 seq.
  62.             with X Mag. x 10 to 5nS/div ▒5%
  63.             Delay Pos 1000:1 with Digital display ▒0.1%
  64. Operating Modes:        A+ Display with main timebase Aonly
  65.             Alt. altern display with intensified A and Del.B
  66.             B display with delayed time B only
  67. Selectable:             Delay time, untriggered or triggered start,slope
  68.             trigger lever,B intensity,vertical A to B trace sep
  69. Bandwidth X-Amplifier:        DC -4MHz (▒3dB). Input X-Amplifier via Ch.1
  70. X-Y Phase Shift:        <3║ below 120kHz
  71. Ramp Output Approx.:        5V positive going
  72. Accessories Supplied:    Line cord, Operators Manual, 2 x Probes-HZ51
  73. Optional Accessories:    50R feedthrough termination HZ22
  74.                         Viewing Hood HZ47,Carry Case HZ96
  75.  
  76. Note:      All Hameg Oscilloscopes have a 2 year warranty